JIXI-H-100A is a trolley-type emergency ventilator manufactured by us.
Now we will introduce how to set PEEP when using the model.
 
The range of PEEP is 0, 5-20hPa.
During exhalation period, the ventilator opens the exhalation channel according to the set PEEP data.
 
Note:
l The lowest PEEP set data should be higher than the present trigger set data, and the highest PEEP set data should be 5hPa lower than the present peak airway pressure.
When adjust PEEP, the tidal volume and oxygen concentration will be changed, and please modify in time.